Beijing Huayi Tongtai Technology Co., Ltd. specializes in the sales of imported analytical and testing instruments. We provide a wide range of products, competitive prices, and excellent after-sales service.
The T2000 and T2100 clamp type grounding resistance testers produced by HT Company in Italy are designed specifically for measuring grounding resistance using the loop method, where the grounding probe can form a loop. The instrument can measure the grounding resistance of a single grounding pile assuming it does not affect each other. The instrument consists of two pliers inside, one for testing current and the other for testing voltage. The T2000 model also performs AC current measurement up to 20A, with a leakage current resolution of 0.05mA, and can set an alarm threshold. The T2100 model is equipped with a half duplex RS-232 interface, which can be connected to the MACROTESTG3 instrument and transmit the data stored on it. The instrument complies with IEC/EN61010-1, CAT III 150V, Very suitable for conducting simple measurements on grounding devices in industrial environments and quickly determining faults.ItalyHT T2000/T2100Clamp on grounding resistance meterProduct Introduction:
T2000/T2100Clamp on grounding resistance tester is an instrument designed to measure the grounding resistance of a complete grounding circuit using the loop method. It is not suitable for testing the grounding resistance of incomplete circuits. When using instruments for this type of measurement, it is assumed that a single grounding device does not affect each other; Therefore, it is commonly used for rapid measurement and qualitative judgment of on-site grounding resistance, which can improve work efficiency. Internally, it is composed of2Composed of pliers, one for testing current and the other for testing voltage.T2000It can also measure AC current up to20AThe leakage current resolution has reached0.05Milliampere and adjustable alarm threshold.
ItalyHT T2000/T2100Product features of clamp type grounding resistance meter:
-Test the grounding resistance of the grounding device using the loop method
-Detecting that there is no connection fault in the grounding pile
-Measure the leakage current of the grounding device
-Set buzzing alarm during testing
-Store test results
-Detecting noise current during testing
-data hold
-Backlight display
-Automatic shutdown
ItalyHT T2000/T2100Product parameters of clamp type grounding resistance meter:
1、 Reference conditions |
|||
parameter |
Reference conditions |
||
reference temperature |
20℃±3℃ |
||
relative humidity |
50%RH±10% |
||
battery voltage |
6.0V±0.5V |
||
external magnetic field |
<30a>30a> |
||
External electric field |
<1v>1v> |
||
Position of pliers |
level |
||
The conductor is located at the position of the pliers |
center |
||
Conductor proximity distance |
>10cm |
||
Secondly, the accuracy of electrical parameters is based on(%reading+This is a parameter based on reference conditions |
|||
grounding resistance measurement |
|||
range |
resolution |
accuracy |
|
0.010~0.099 |
0.001 |
±(1%rdg+0.01Ω) |
|
0.10~0.99 |
0.01 |
±(1%rdg+0.01Ω) |
|
1.0~49.9 |
0.1 |
±(1.5%rdg+0.01Ω) |
|
50.0~99.5 |
0.5 |
±(2.0%rdg+0.5Ω) |
|
100~199 |
1 |
±(3%rdg+1Ω) |
|
200~395 |
5 |
±(6%rdg+5Ω) |
|
400~590 |
10 |
±(10%rdg+10Ω) |
|
600~1000 |
20 |
±(20%rdg+20Ω) |
|
Frequency range for resistance measurement:>1KHz,Measure the resistance alarm threshold: 1-999Ω |
|||
Measurement of current (true effective value) |
|||
range |
resolution |
accuracy |
|
0.00mA~80.00mA |
0.05mA |
±(2.5%rdg+1mA) |
|
80.00mA~650.0mA |
0.5mA |
±(2.5%rdg+2mA) |
|
650.0mA~999.5mA |
1.5mA |
±(2.5%rdg+10mA) |
|
1.000A~3.99a |
0.00a |
±(2.5%rdg+0.1A) |
|
4.00A~20.00A |
0.01A |
±(2.5%rdg+0.a) |
|
Frequency range for current measurement: 50-60Hz,Set current measurement threshold: 1mA-999mA, Waveform distortion: CF<>1.414 andCF<>add1.5%The error. |
|||
3、 Basic specifications |
|||
mechanical properties |
|||
size |
293×90×66mm |
||
Weight (including battery) |
1320can |
||
Large jaw diameter |
32mm |
||
working conditions |
|||
Reference temperature: |
20℃±3℃ |
||
Working temperature: |
0-40℃ |
||
Working humidity: |
10%RH - 90%RH |
||
Reference standards: |
|||
Following standards: |
IEC/EN61010-1,IEC/EN61010-2-032 |
||
isolation: |
Double insulated |
||
pollution degree: |
2 |
||
security level: |
CAT III 150Vto ground,very big20A. |
||
other: |
|
||
Type: |
4×1.5V LR6 AA MN1500 |
||
Automatic shutdown: |
stop5 Minutes later |
||
storage capacity: |
99 A data |
||
Display Features: |
4LCDDot matrix display with backlight |
||
Display update frequency: |
1 order/second |