one overview
high speedCentrifuge 5/7ml * 10Separation technology is a new technology developed based on the state of particles in a practical centrifugal environment. Particles of different densities, sizes, or shapes settle in different centrifugal situations, so a roughly spherical non-uniform mixture can be separated by centrifugation. Centrifuges are used to further study biochemistry and separate large amounts of substances. For example, collecting cells, separating plasma, and isolating from these pre purified preparationsDNAAnd protein molecules can be used to isolate viruses as well as large-scale Escherichia coli, strict cellular components, nuclear protein particles, etc.
This machine adopts a brushless motor, microcomputer control, and hasRCFAutomatic calculation and setting. It has the advantages of multiple program protections, easy operation, and maintenance free.
two High speed centrifuge 5/7ml * 10Main technical parameters
rotational speed |
16000rpm |
time range |
0-99mins |
maximum centrifugal force |
19200×g |
Fuse specifications |
10A |
Maximum capacity |
400ml |
Power supply |
220V 50Hz |
Motor Power |
450W |
Weight |
25kg |
External dimensions |
490×400×370(L×W×H)mm |
Equipped with rotor
Rotor number |
Rotor model |
speedr/min |
centrifugal force×g |
rotor type |
1 |
12×1.5/2.0ml |
16000 |
19200 |
ANGLE ROTOR |
2 |
24×1.5/2.0ml |
13000 |
16060 |
ANGLE ROTOR |
3 |
48×0.5ml |
13000 |
14740 |
ANGLE ROTOR |
4 |
10×5/7ml |
13000 |
11500 |
ANGLE ROTOR |
5 |
6×50ml |
12000 |
14800 |
ANGLE ROTOR |
6 |
12×10ml |
11000 |
14610 |
ANGLE ROTOR |
7 |
12×15ml |
10000 |
10290 |
ANGLE ROTOR |
8 |
4×100ml |
10000 |
9690 |
ANGLE ROTOR |
9 |
2×48hole |
3000 |
1400 |
Enzyme linked immunosorbent assay plate horizontal rotor |
Usage
1Place the centrifuge on a sturdy and horizontal platform, ensuring that all four legs are evenly distributed
Apply force, then gently shake with your hand to check if the centrifuge is placed smoothly.
2Open the door cover, gently place the rotor onto the motor shaft, tighten the nut to tightly connect the rotor to the motor shaft (be careful and gentle to avoid significant left and right shaking and up and down pressure to avoid damaging the flexible support of the motor), and then gently rotate the rotor body by hand to ensure flexible rotation.
3Place the centrifuge tube into the rotor test tube hole, ensuring that it is fully inserted and the test solution inside the tube is evenly distributed to ensure balanced operation of the instrument.
4Gently close the door cover and check if it is locked properly (after closing, gently lift the door cover with force, but it cannot be opened).
5Plug in the power socket and press the power switch.
6Set rotor number, speed, time