Technical parameters of high-altitude low-pressure simulation test chamber and programmable high and low temperature low-pressure test chamber
1、 Low pressureThe experiment shall be conducted according to the following steps:
Connect the battery according to4.5.1After being fully charged according to the prescribed test method, place the battery in20℃±5Place it in a vacuum test chamber at ℃. Vacuumize to reduce the pressure inside the box to11.6kpa(Simulating altitude)15240m)And continue to do so6h.
The battery should not ignite, explode, or leak.
2、 Main technical parameters
|
rule grid |
model |
CZ-DQ-01 |
CZ-DQ-01T |
CZ-DQ-02T |
|
|
internal volume |
175L |
||||
|
studio dimensions |
W500×H700×D500mm |
W500×H700×D500mm |
W500×H700×D500mm |
||
|
Overall dimensions |
About W880×D730×H1680mm |
About W980×D830×H1680mm |
About W980×D830×H1680mm |
||
|
box body knot constitute |
structure |
monolithic |
|||
|
outer box material |
High quality thickened cold-rolled steel plate (gray+blue) baked paint |
||||
|
Inner box material |
High quality thickened stainless steel |
||||
|
observation window |
Tempered glass observation window (350×400mm double-layer explosion-proof glass) |
||||
|
LED lighting |
|||||
|
main want skill technique point mark |
Control method |
PLC touch screen |
|||
|
Vacuum pressure range |
0-99kpa (adjustable) |
||||
|
bias |
±0.5kpa |
||||
|
precision |
0.1 |
||||
|
volatility |
±0.5 |
||||
|
holding time |
0-99 hours, 99 minutes, and 60 seconds |
||||
|
temperature range |
room temperature |
RT+10℃~+150℃ |
-40℃~+150℃ |
||
|
heating rate |
/ |
RT→+150℃ ≤1℃/min |
RT→+150℃ ≤1℃/min |
||
|
temperature resolution |
/ |
0.1℃ |
|||
|
temperature fluctuation |
/ |
±0.5℃ |
|||
|
temperature deviation |
/ |
≤2.0℃ |
|||
|
Temperature Uniformity |
/ |
≤2.0℃ |
|||
|
Standard Configuration |
Host and sample rack 2Insulation epoxy resin sample partition |
||||
|
power supply voltage |