Hash 2100Q portable turbidity meter
The 2100Q turbidity meter is based on the world's classic 2100P portable turbidity meter optical system, using Hash Company's ratio measurement. The dual detector optical system can compensate for the chromaticity, light fluctuations, and stray light of the sample. For most samples, even under harsh on-site environmental conditions, laboratory level measurement performance can still be obtained.
Hash 2100Q?Features and advantages:
Accurate testing of optical systems can be conducted on-site For rapidly settling samples, the measurement is accurate. ● Easy calibration and verification ● Convenient data transmission ● Convenient data storage function There are two models that can meet your needs: 2100Q turbidity meter - compliant with USEPA method 180. 2100Q IS turbidity meter - compliant with ISO7027;
|
Technical parameters of portable turbidity meter:
|
Compliant with standards: |
2100Q: Compliant with USEPA180.1; 2100Q IS: Compliant with: ISO 7027 |
|
|
Light source: |
2100Q: Tungsten lamp; 2100Q IS: LED@860nm |
|
|
Range: |
0-1000NTU |
|
|
Accuracy: |
0-1000NTU: ± 2% of reading+light scattering |
|
|
Repeatability: |
± 1% of the reading or 0.01NTU, whichever is greater |
|
|
resolving power: |
Zui low 0.01 NTU |
|
|
Scattered light: |
<0.02NTU(FNU) |
|
|
Signal averaging mode: |
Optional to open or close |
|
|
Detector: |
silicon photocell |
|
|
Reading mode: |
User selectable: Normal mode, Signal averaging mode, Quick setting turbidity mode |
|
|
Data storage: |
500 items |
|
|
Power requirements: |
110-230VAC, 50/60Hz (with power supply or USB+power component) |
|
|
|
Four AA alkaline batteries or rechargeable batteries (for use with USB+power components) |
|
|
Work environment: |
0-50℃; 0-90% @ 30 ℃ relative humidity; 0-80% @ 40 ℃, 0-70% @ 50 ℃ without condensation |
|
|
Save environment: |
-40-60 ℃, instrument only |
|
|
Interface: |
Optional USB |
|
|
Instrument protection level: |
IP67 |
|
|
Zui small sampling volume: |
10ml |
|
|
Sample colorimetric tube: |
60 × 25mm, borosilicate glass and screw cap |
|
|
Size: |
229×107×77mm |
|
|
Weight: |
Without battery: 527g; Containing AA alkaline battery: 618g |
|
|
Optional languages: |
English, French, German, Italian, Chinese, etc |
|
Choose accessories: |
|
||
LZV813: USB+power component, including power adapter, USB cable, operation card |
|
||
LZV804: Power components, including power adapter and operation card |
|
||
2971304: NiMH batteries, 4 cells per pack |
|
||
4397500: Degassing kit |
|
||
4397510: Sample filtration+degassing kit |
|
||
2971210: StablCal calibration solution, 100ml, used for 2100Q |
|
||
2971200: TablCal calibration solution, 500ml, used for 2100Q |
|
||
2464105: Gelex Secondary Standard Liquid |
|
Standard configuration:
Hash Company's portable turbidity meter is equipped with 4 AA alkaline batteries, a portable case containing (20100800NTU) StablCal primary calibration solution in a 1-foot sealed bottle, 10NTU primary validation standard, 6 sample cells with screw caps, instrument operation manual (printed and CD), quick start guide, silicone oil, and oilcloth.
Common troubleshooting:
Error/Warning |
describe |
Solution |
Close the lid and press' Read ' |
Lid open or lid detection failed. |
Ensure that the lid is closed during the reading and re reading period |
Low battery! |
It's due to low battery power |
• Install a new battery If using rechargeable batteries, connect the USB/power module |
ADC malfunction! |
Hardware error causing reading failure |
Readjust again |
The detector signal is too low! |
Insufficient lighting on the 180 ° detector |
Check if the optical path is blocked Please check the light |
Out of range! |
High turbidity - may be due to using only RapidCal ™ Calibration caused. |
Calibration upper limit. Dilute the sample |
Below the range! |
The measured absorbance is below the calibration range |
• Repeat calibration |
Please check the lights! |
The signals from the 90 ° and 180 ° detectors are too low |
2100Q: There is a problem with the light. Replace the light (please refer to the instruction manual for bulb replacement). 2100Qis: Please communicate with the department. |
The temperature is too high! Turn off the instrument |
The temperature has exceeded the instrument limit (>60 ° C or>140 ° F) |
Turn off the instrument panel and let it cool down |
RST: The average value! |
The solid substance precipitates too slowly, and the reading mode is not suitable for this sample |
Please select the 'Normal' or 'Signal Average' reading mode |