Technical Parameter
model |
CR219-4 |
Control range |
50~180℃ |
Dissolving temperature |
50~180℃ |
Temp Accuracy |
±0.1℃ |
Temperature tolerance |
±1℃ |
Sample processing quantity |
Can process 4 water samples simultaneously |
Power Supply |
DC 12V/10A |
Size |
110*170*125mm |
Weight |
918g |
precautions
lThe digestion temperature should be controlled between 50-180 ℃
lThe uniformity of the temperature field of the digestion device should not be greater than 3 ℃
lThere should be no water or other foreign objects inside the heating hole
lEnvironmental temperature: (0~40℃)
lRelative humidity: not exceeding 85%
lPower supply: DC 12V/10A
lThere are no corrosive gases present in the surrounding air
lVibration that does not affect performance; Electromagnetic field interference that does not affect performance other than the geomagnetic field
lAfter digestion is complete, take out the test tube and place it on the test tube rack. Let it cool naturally to below 100 ℃ and shake the reagents in the test tube evenly. Then, place the test tube on the test tube rack and let it cool naturally to room temperature for testing
lNote: When cooled to room temperature, shake the test tube and directly take the upper clear liquid for testing
Standard configuration
CR219-4 host, power cord, instruction manual, warranty card/certificate of conformity.