-Product Introduction:
DRB200 COD Dissolver/Dissolver is a recently developed digestion device by Hash Company, which can perform COD TOC、 The digestion of water samples such as total phosphorus and total nitrogen. The digestion time and temperature can be selected within the range of 0-480 minutes and 37-165 ℃, respectively. In order to meet a wider range of usage needs, a heating block or two independent heating blocks can be configured in the DRB200 COD digester, which can perform digestion with two independent heating control modules under different/the same temperature and different/the same time setting conditions. The heating block comes in three sizes, suitable for pre packaged TNT plus reagents with barcodes (13mm), regular TNT pre packaged tubes (16mm), and pre packaged tubes for metal digestion pretreatment (20mm).
-Working principle:
The DRB200COD digester/analyzer can heat and digest COD for different pre made reagents through a heating module and preset heating program TOC、 Total phosphorus, total nitrogen and other water samples.
-Application industry:
The DRB200 COD digester/analyzer can be used in industrial, municipal, environmental protection, university research and other fields for COD TOC、 Heating digestion of water samples containing total phosphorus, total nitrogen, etc.
-Instrument features:
The digestion temperature and time can be selected, with an adjustable temperature range of 37 ° C to 165 ° C and an adjustable time range of about 1-480 minutes.
The fully insulated module design allows for simultaneous digestion reactions of two independent heating control modules under different/same temperature and time settings.
The DRB200 digester is equipped with multiple heating aperture options, which can be used for COD TOC、 The digestion of water samples such as total phosphorus and total nitrogen.
The DRB200 disintegrator has a built-in stored program that allows users to select with just one click, eliminating the hassle of manual settings and potential errors.
Equipped with a digital timer, automatic cut-off and alarm functions, safe and reliable.
Fast heating and temperature stability better than ± 1 ° C.
The DRB200 digester adopts an isolated locking design, equipped with a transparent protective cover and an anti overheating safety device.
TECHNICAL INDEX | |
---|---|
specifications | detailed information |
size | 25 x 14.5 x 31 centimeters (9.8 x 5.7 x 12.2 inches) |
weight | Single heating zone: 2 kilograms (4.4 pounds) |
Dual heating zone: 2.8 kilograms (6.17 pounds) | |
Power requirements | 100-240V,+ 5%/-15%, 50/60Hz, protection level I |
power input | Single heating zone: 115V: 300VA: 230V: 450VA |
Dual heating zone: 115V: 600VA: 230V: 900VA | |
Storage temperature | -40 to+60 ℃ (-40 to+140 ° F) |
Operating temperature | 10 to 45 ℃ (50 to 113 ° F) |
Operating humidity | 90% (no condensation) |
Instrument range | Temperature: 37 to 165 ℃ (98.6 to 329 ° F) |
Timer: Approximately 0 to 480 minutes | |
temperature stability | ±2℃(±3.5°F) |
Heating speed | It takes about 10 minutes to go from 20 to 150 ℃ (68 to 302 ° F) |
authentication | CE、GS、 And cTUVs |
Accessory information | |
---|---|
28958-05 | Can convert 16mm pores to 13mm pores for digesting TNTPlus pre made reagents |
HHA155 | Can convert 20mm holes to 16mm holes |
Ordering Guide | ||||
---|---|---|---|---|
Order number | Number of heating blocks | The number of 13mm holes | Number of 16mm holes | Number of 20mm holes |
LTG082.03.30003 | 1 | / | 9 | 2 |
LTG082.03.40003 | 1 | / | 15 | / |
LTG082.03.42003 | 2 | / | 21 | 4 |
LTG082.03.44003 | 2 | / | 30 | / |
DRB200-05 | 1 | 9 | / | 2 |
DRB200-06 | 2 | 21 | / | 4 |
DRB200-07 | 2 | 30 | / | / |
DRB200-08 | 2 | 12 | / | 8 |