Product Introduction:
MG268S-H type is a high gloss board; MG268S-M type is a medium gloss board; MG268S-L is a low gloss board
Specification: 45mm × 85mm
Annual variation ≤± 0.5 gloss unit
The working face of MG268S-H type has an interference aperture of no more than 15 apertures and a local aperture of no more than 3 apertures
effect:
Used for the calibration of glossiness meters.
matters needing attention:
During the operation, the entire set of instruments should be kept clean. Otherwise, it will cause measurement errors and inaccurate calibration
If there is fog, dust, or stains on the surface of the work board, please use lens paper or a soft cloth dipped in anhydrous ethanol to wipe it clean
The measuring port of the instrument should be in contact with the measured surface to avoid external light leakage. If necessary, black cloth or other objects can be used to block the light around the measuring port
Suggest avoiding strong light environment measurement
Calibration, measurement, or verification operations should be carried out at the same ambient temperature, otherwise a large temperature difference in the environment will cause digital offset
If the MG6 series glossiness meter is not used for a long time, the battery should be removed to avoid corrosion and damage to the instrument
If the WGG60 series glossiness meter is not used for a long time, the internal battery pack should be recharged with an appropriate amount of electricity every month.