- Product Name: Gas permeability tester - pressure difference method
- Product Code: BSG-11-A
BSG-11-A
Function
Gas permeability performance test.
Suitable for testing the permeability of O2, CO2, N2 and other non-toxic gases in plastic films, composite films, high barrier materials, sheets, metal foils, rubber and other materials.

Working principle
Differential pressure method.
Place the pre-set sample between the high-pressure chamber and the low-pressure chamber, seal it tightly, and then evacuate both chambers simultaneously; After vacuuming for a certain period of time and the vacuum degree drops to the required value, close the low-pressure chamber, fill the high-pressure chamber with test gas, and adjust the pressure of the high-pressure chamber to maintain a constant pressure difference on both sides of the sample; Gas permeates from the high-pressure side to the low-pressure side of the sample under the action of pressure difference; Accurately measure the pressure changes in the low-pressure chamber and calculate the gas permeability parameters of the sample.
Compliant with standards:
YBB 00082003、GB/T 1038、ASTM D1434、ISO 2556、ISO 15105-1、JIS K7126-A。
FEATURES
- Imported high-precision vacuum sensors and pressure sensors with high testing accuracy;
- Semiconductor cold and hot bidirectional temperature control, parallel connection, high reliability;
- Precision valve pipeline components, thoroughly sealed, achieve high-speed vacuum, thoroughly desorb, and reduce testing errors;
- Sample anti leakage design to reduce sample installation leakage;
- Precise pressure control, maintaining a wide range of pressure differences between high and low pressure chambers;
- The patented baseline technology removes system background leaks and reduces testing errors;
- Intelligent automatic: power on self-test to avoid continuing testing under fault conditions; One click start, fully automated testing execution.
- Data recording: graphical, full process, all element recording, no loss of power-off data.
- Data Security: The optional "GMP Computerized System" software module has functions such as user management, permission management, and data audit tracking.
- Working environment: Indoor. No need for constant temperature and humidity environment (reducing usage costs), test data is not affected by environmental temperature and humidity.
TECHNICAL INDEX
|
name |
parameter |
name |
parameter |
|
test error |
0.01 cm3/m2•day•0.1MPa |
Correction method |
Standard film |
|
Vacuum error |
0.1Pa |
Vacuum range |
1333.33 Pa |
|
Vacuum degree |
<5 Pa |
Vacuum efficiency |
Within 20 minutes, below 27Pa |
|
temperature range |
15℃~50℃ |
Temperature control error |
±0.1℃ |
|
sample thickness |
≤3mm |
test area |
50cm2rotundity |
|
test gas |
O2、 N2、 CO2Non toxic gas |
test pressure |
0.1~0.2 MPa |
|
Gas interface |
1/8” |
Mpa |
0.1~0.8 MPa |
|
Power source |
AC220V 50Hz |
Power supply power |
<1500 W |
|
Model differentiation |
Model number |
||
|
BSG-11A |
|||
|
measuring range
|
0.1-50,000 |
||
|
Sample quantity |
1 |
||
|
Number of vacuum sensors |
1 |
||
|
Test mode |
Single cavity independent |
||
|
Host size (L × B × H) |
585×640×380mm |
||
|
Weight |
40Kg |
||
Standard configuration
Test host, vacuum pump, testing software, sampler, DELL computer (BSG-11 model not available), vacuum bellows, gas cylinder pressure reducing valve and fittings, sealing grease.
Select accessoriesContainer testing fixture, humidity control unit.
Self provided partsTesting gases and cylinders, testing computer (BSG-11 type).