Four probe resistivity tester for powder testing method Four probe resistivity tester for powder testing method
Product Name: Powder resistivity tester (four probe method)
Product model: FTDZS-I
Reference standard: YS T 587.6-2006 Testing method for calcined petroleum coke for carbon anodes Part 6: Determination of powder resistivity. pdf
GB/T 24525-2009 Determination of electrical resistivity of carbon materials
GBT 30835-2014 "Carbon composite lithium iron phosphate cathode materials for lithium-ion batteries" specifies the requirements for the four probe method used in the arbitration method for the determination of powder conductivity
Product Overview:
The powder resistivity tester is mainly used as a specialized instrument for measuring the resistivity of powder materials
Due to the different densities of powder materials, the data obtained from testing under loose and compacted density conditions are different. Therefore, testing powder electricity
Resistance requires testing to be conducted under specified pressure conditions Facilitating effective data testing and comparison
The instrument consists of two main parts: the host and the testing rack. The host includes a highly sensitive DC digital voltmeter and a highly stable DC constant current source for measurement
The experimental setup is operated and set through a touch screen, with a reasonable page layout and user-friendly design, and the test results can be printed.
The instrument has the characteristics of high measurement accuracy, good stability, compact structure, and easy use, which meet the requirements of international and national standards.
The instrument is suitable for industries such as conductor powder, semiconductor powder, and carbon materials for testing and inspection.
Main technical specifications of the instrument:
1、 Measurement range:
When measuring the resistivity of a semiconductor: resistivity 10-5-105 Ω - cm; resolution 10-6 Ω - cm
When measuring the resistance of other wires and cables: resistance 10-4-2X105 Ω, resolution 1u Ω
2、 Voltage measurement:
1. 量程 2 mV、20 mV、200 mV、2V
2. Measurement error 2mA ± (0.5% reading+8 words); 20mV-2V gear ± (0.5% reading+2 words)
3. Display 4 1/2 digit numbers from 0 to 19999 with polarity and overload automatic display, decimal point and unit automatic display.
3、 Constant current source:
1. Current output: 1 μ A, 10 μ A, 100 μ A, 1mA, 10mA, 100mA
2. Current error: ± (0.5% reading+2 words)
4、 Measurement electrode:
1. Measuring electrode: four probe
2. Probe spacing: 1MM
3. Electrode Material: Tungsten Rod
4. Tablet diameter: 2cm Area 3.142cm ^ 2
5. Powder sample size: 2.501g-2.5099g
5、 Loading device:
1. Loading electrode 1: manual loading
2. Loading electrode 2: automatic loading
3. Sensor range: 10KN