FP640 flame photometer (specifically designed for the cement industry)
(Mainly used to detect the content of potassium oxide and sodium oxide in cement. Refer to GB/T 176-2008 Cement Chemical Analysis Method)
Main performance parameters
·Potassium and Sodium
·Dual channel proportional value display
·Readings can display decimal points
·Rotary knob operation
·Pre selected flame size
·Equipped with flameout protection device
·Repeatability: ≤ 3%
·Sample suction and spray volume:< 6mL/min
·Response time:< 8s
·Spectral splitting method: interference filter
·Display mode: LED LCD screen
·Stability: ≤ 3% within 15 seconds; ≤ 5% within 5 minutes
·Detection limit: K:≤0.004mmol/L; Na:≤0.008mmol/L
·Linear error: K:≤0.005mmol/L ; Na:≤0.03mmol/L
·Test scope: K:0~100ug/ml; Na:0~160ug/ml
·Transmittance characteristics of filter: potassium: absolute value of peak wavelength error ≤ 7nm, sodium: absolute value of peak wavelength error ≤ 5nm
Potassium: Half width ≤ 15nm Sodium: Half width ≤ 15nm
Potassium: background transmittance ≤ 0.5% Sodium: background transmittance ≤ 0.5%
application area
·Cement · Third party testing · Professional colleges