VIP member
ET8500 microcomputer urea CO (NH2) 2 concentration analyzer
ET8500 urea CO (NH2) 2 concentration analyzer
Product details
ET8500 urea CO (NH2) 2 concentration analyzer
Range: 0.1 to 2.5 mg/L (NH2) 2CO, 0.2 to 5 mg/L (NH2) 2CO measurement, infrared data transmission function, 16 sets of data storage, can be accessed at any time, GLP management function, factory calibration and user-defined calibration dual modes, waterproof design, suitable for rapid sample analysis and measurement in laboratories and on-site
Standard configuration● Exquisite and portable, easy to operate, precise and fast measurement
● Background light display screen, GLP function, automatic shutdown setting
16 sets of data storage, which can be accessed and read at any time, making it convenient for on-site data organization and recording
Infrared transmission design, IRiM infrared transmitter (ET214050) and computer connection
● Factory and user-defined dual calibration function, automatically remembers the last zero calibration when the instrument is turned off, eliminating the need to calibrate the instrument zero every time it is measured, ensuring accurate and reliable measurement
Waterproof sample pool design, prefabricated supporting reagents, more reasonable operation and maintenance
● Applied to water treatment, wastewater/drinking water, cooling water/boiler water, swimming pool water, etc
measuring principle
measuring principle
Low energy LED is used to provide stable light source, and the measurement curve is built-in. The reagent and sample undergo color reaction, and the absorbance of the sample at a specific wavelength is converted into the concentration value of the parameter to be measured, which is displayed on the LCD screen
Convenient and efficient
Using a light source (LED) to provide a stable light source with low energy consumption. User friendly interface, easy and fast operation. Equipped with dual calibration functions of factory calibration and user-defined calibration, ensuring accurate, stable, and reproducible measurements.
Using a light source (LED) to provide a stable light source with low energy consumption. User friendly interface, easy and fast operation. Equipped with dual calibration functions of factory calibration and user-defined calibration, ensuring accurate, stable, and reproducible measurements.
| Technical specifications for ET8500 urea CO (NH2) 2 concentration analyzer | ||
| Host model | MD100 | |
| range | 0.01-2.5mg/L CO(NH2)2 | 0.2-5mg/L CO(NH2)2 |
| accuracy | ±0.02mg/L Urea | ±0.4mg/L Urea |
| resolution | 0.1mg/L | 0.1mg/L |
| Reagent type | ET459300 reagent (15ml water solution), ET459400 reagent (10ml water solution), ET517611 reagent (tablet 100) | |
| measuring method | Urease/indigo colorimetric method | |
| optical system | 610nm, Waterproof sample pool with temperature compensated LED optical system, wavelength accuracy ± 1nm | |
| Photometric index | Resolution: 0.01A, Accuracy: 3% F.S | |
| Calibration Mode | Automatic zero point calibration, using SMD microprocessing technology, with factory calibration and user-defined calibration modes | |
| data management | 16 sets of measurement data are automatically stored using the IRiM infrared data transmission module | |
| Power supply mode | 4X1.5AAA battery with automatic shutdown and power-saving function (10 minutes) | |
| Usage environment | 5-40 ° C; RH: 30-90% non condensing | |
| Dimension | 155×75×35mm/260g | |
MD100 host, standard ET459300 reagent (water solution, 15ml), ET459400 reagent (water solution, 10ml), ET517611 (composed of ET512580 reagent and ET512590 reagent) reagent (tablet, 100), ET197620 (Ø 24mm) dedicated glass colorimetric dish (2 pieces), ET197635 colorimetric dish cleaning cloth (2 pieces), Chinese and English user manual, dedicated carrying case
Online inquiry