According to Lambert Beer's law, infrared photometry is used to determine petroleum, animal and vegetable oils in water, and oil fume in air at wave numbers of 3030, 2960, and 2930cm-1The product meets the following standards for measuring substances with characteristic absorption:
*HJ-637-2018 Water Quality - Determination Standards for Petroleum, Animal and Vegetable Oils
*GB18483-2001 Emission Standard for Cooking Fumes in the Catering Industry
Functional Features
◆ Stable dimming signal, fast response time, and long lifespan of the light source
Good reproducibility, measurement error not exceeding ± 2% (concentration 40mg/L)
Zero point memory, one-time zero adjustment for multiple measurements
High baseline stability, absorbance drift less than 0.005AU/30min
◆ Built in 10 standard curve calibration functions
◆ Dual system analysis function with carbon tetrachloride and tetrachloroethylene
◆ 8-inch LCD display, USB 2.0 interface for convenient data communication
Quick calibration function, only one concentration standard sample is needed to complete
◆ It has functions such as input, storage, scanning analysis, measurement, calibration, data processing, and printing of sample information
TECHNICAL INDEX
Measurement range: 0-1000 mg/L
Detection limit: 0.2mg/L (carbon tetrachloride extract)
0.4mg/L (tetrachloroethylene extract)
Reproducibility: RSD ≤ 1% (40mg/L oil)
Linear coefficient: ≥ zero point nine nine nine
Wave number scanning range: 3400-2400 cm-1
Wave number accuracy: ±2 cm-1
Wave number reproducibility: 1cm-1
Sample measurement time: 35 seconds
Reading method: concentration (mg/L), absorbance (abs), transmittance (T%)
Working temperature: 5~35°C
External output: USB interface
External output: USB interface
Working power supply: 220±10%VAC,50Hz
External dimensions: 50cmX38cmX22cm
Instrument weight: 15kg
Instrument usage
As a specialized instrument using the infrared spectroscopy method developed by the Ministry of Environmental Protection, it can be used to determine the content of various oil products in surface water, groundwater, domestic sewage, and industrial wastewater, as well as to measure the concentration of cooking fumes in the catering industry. Its measurement results are comparable.
As a universal infrared spectrophotometer, it can scan and analyze the structure of substances and perform qualitative and quantitative analysis of various organic compounds.
Environmentally friendly tetrachloroethylene is recommended for solvents and extractants, and other organic solvents that have been tested and approved are also acceptable.
application area
Environmental monitoring, hydrology, surface water, petrochemicals, oil fields
Ports, mining machinery, electronic tap water, sewage treatment plants
Disease control, health, schools
Instrument accessories
windows operating system
Automatic liquid-liquid extractor
Exquisite tetrachloroethylene that meets the requirements of the method
Two specifications of quartz colorimetric dishes