type number (Manual type, automatic type) |
OptiCenteric |
OptiCentric L |
OptiCentric MOT |
OptiCentric MOT 2R |
Sample diameter range Diameter of air flotation platform |
225 |
225 |
225/600 100/300 |
225/400/600 100/300 |
Standard measurement range (Focal length, curvature radius) |
±5 …±250mm |
±5 …±400mm |
±5 …±450mm |
±5 …±450mm |
Focal length adjustment mode |
Linear guide rail lifting Manually adjust the focus5um Height adjustment stroke250 |
<-- Manually adjust the focus5um |
Electric lifting,Auto focus Electric focusing step0.5um |
<-- Electric focusing step0.5um |
size(mm) weight(kg) |
355 x 455 x 970 35 |
355 x 490 x 1000 35 |
550 x 750 x 1733 60 |
750 x 1150 x 2000 80 |
Scalable measurement range Measurement accuracy and mode |
Scalable to±2000 mm(Focal length, curvature radius); Ensure measurement accuracy:+/- 0.2micronperhaps+/-2second; Measurement mode-Transmission and reflection methods, axial measurement |
|||
Measurement parameters And expansion |
Measurement of spherical center deviation using reflection and transmission methods; Centering and bonding of lenses; The center deviation of the composite lens system. Optical disc and optical drive components:CDCalibration of optical reading head; The oscillation degree of glass disk rotation.Wedge angle of planar optical components. BFLCurvature radius, lens thickness, non spherical center deviation, planar component angle, etc. |
|||
Other parameters |
Collimatorfocal length150mm, optical aperture30mmMeasure the focal length of the measuring head200mmMeasure the wavelength of green light; Precision rotatable fixture; Tiltable dual axis adjustment tableTRT200Equipped with a micro micrometer for precision adjustment2second/1umAxial and radial accuracy of air flotation platform0.05umIdeal temperature:20-22Degree(0-40)Humidity30-70% ;*Atmospheric supply pressure:1 MPa (10 bar) |