Temperature range: Room temperature to 500 ℃
Transparent hole: 2.5 mm in diameter
Heating method: double-sided heating
Heating on/off: fixed gap or adjustable gap
Window to sample distance: 7 mm
Maximum heating rate: 80 ℃/min
Temperature display and control accuracy: 0.1 ℃/0.3 ℃
Temperature control: regulator accuracy level 0.2, PID artificial intelligence control, 50 segment program
Temperature control software: interface switching between Chinese and English, real-time control, multi-point temperature correction, adjustable temperature speed, temperature control range, etc., process curve display and recording
Experimental environment: The chamber is sealed and the atmosphere can be passed through
Cooling method: circulating water cooling
Complete configuration: hot table, circulating cooling water pump, hot table temperature controller, temperature control software, foil maker, etc