High precision turbidity analyzer, user-friendly interface design, advanced optical system, ensuring accurate analysis and measurement. Equipped with a dedicated calibration group, quick calibration can be performed at any time. The instrument has GLP good laboratory management function, and you can check various relevant information of the last calibration, such as calibration point, calibration time, etc. User friendly interface design, during the operation process, corresponding prompt information will be displayed according to the operating conditions, providing you with step-by-step operation guidance. The instrument has excellent storage function, with 200 built-in data storage spaces and the ability to access stored data at any time. At the same time, if you choose data analysis software and data cables, you can also transfer stored data to a PC through a USB data interface for further data analysis.
Product parameters
model |
DYT-A11N |
Turbidity range |
0.00 to 9.99; 10.0 to 99.9; 100 to 1000 FNU ; |
accuracy |
± 1% of reading+0.02 NTU; ± 5% of the reading (reading greater than 1000 NTU) |
reproducibility |
± 1% of reading+0.02NTU |
Measurement mode |
Ordinary measurement, continuous measurement, average measurement |
optical system |
Infrared LED light source, silicon photovoltaic cell |
Calibration method |
Turbidity up to 5 calibration points< 0.1,15,100,750,2000 NTU; |
Measurement standards |
ISO 7027 |
Data Storage |
200 sets of measurement data, USB data interface connected to computer (data cable and software need to be purchased) |
Power supply mode |
230 V/50 Hz,20 W |
Usage environment |
0 to 50°C (32 to 122°F); RH max 95% non condensing |
Dimension |
230 x 200 x 145 mm; 2.5 Kg |
DY 92000 | Windows Data Processing Software |
|
DY 920013 | USB data transmission cable |
|
DY 88703-6 | Turbidity Calibration Group - (0.1, 15, 100, 750, 2000NTU) |
|
DY93703-50 | Glass colorimetric dish cleaning solution, 500 mL |
|
DY731318 | Color dish cleaning cloth, 4 pieces per group |
|
DY 731333N | Glass colorimetric dish (cup+lid), packaging: 4 pieces/set |
|
DY731342 | Liquid extractor, standard: 2000 µ l, standard 2000 µ l dedicated liquid extractor head (4 pieces) |
|
DY731352 |
2000 µ l dedicated liquid collection head (4 pieces/group) |
|