●Compliant with environmental protection industry standardsHJ/T 101-2019Technical Requirements and Testing Methods for Ammonia Nitrogen Water Quality Automatic Analyzer;
●Using ammonia sensitive electrode method to determine ammonia nitrogen in water,No need to use mercury containing reagents such as Nessler's reagentEnvironmentally friendly;
●use7Inch color LCD touch screen, Chinese interface, easy to operate;
●Equipped with a constant temperature control device to prevent sample temperature fluctuations from affecting measurement accuracy;
●Support automatic calibration/Measurement, calibration can be freely set/Measure the interval period and support automatic cleaning;
●Support custom settings for detection parameters;
●Support automatic saving of measurement data and calibration data, and support saving and viewing of massive historical data;
●Support alarm function, can set upper and lower limit concentration for alarm;
●Support power-off protection and abnormal reset function;
●Having(4~20)mA DCIsolation signal output, output upper limit and output lower limit can be set arbitrarily;
●haveRS-232perhapsRS-485Digital interface(2select1)Supports data export and remote communication.
model Technical Parameter |
DWG-8002A |
measuring method |
Ammonia sensitive electrode method |
measuring range |
(0.05~1000.00)mg/L |
resolution |
0.01mg/L |
Indication error (Percentage of test solution concentration relative to the upper limit of the basic detection range (10mg/L)) |
20% ±8% 50% ±5% 80% ±3% |
Comparison of actual water samples |
±0.2mg/L(≤2.00mg/L) ±10%(>2.00mg/L) |
repeatability |
≤2% |
stability |
±10%/24h |
zero drift |
≤0.02 mg/L |
Span Drift |
≤1% |
90% response time |
≤6min |
power supply |
AC:(220±22)V,(50±0.5)Hz |
Size (mm), weight(kg) |
550×400×1680,60 |