Shanghai Shouli Industrial Co., Ltd
Home>Products>DUALSCOPE MPOR Coating Thickness Measuring Instrument
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
    18930767719
  • Address
    Room 307, Huacheng Plaza, 1471 Gumei Road, Minhang District, Shanghai
Contact Now
DUALSCOPE MPOR Coating Thickness Measuring Instrument
DUALSCOPE MPOR is mainly used in the automotive and shipbuilding industries, paint and varnish surface treatment, electroplating industry, aerospace i
Product details

Product Introduction

summary:

DUAL SCOPE MPORNot only is it one of the smallest coating thickness gauges, but it is also equipped with: double-sided backlightLCDDisplay, large display panels and set-top display panels can safely and conveniently see the readings of the measured coating thickness from any angle; The built-in standard radio transmitter can directly transmit readings to the computer online or offline, with a wireless distance of10-20m33-66feet

The measured thickness includes:

  • Utilize complianceDIN EN ISO 2360The standard eddy current method can measure the thickness of paint, varnish, fiber, or oxide film coatings on non-magnetic metal materials such as aluminum, copper, brass, and stainless steel;
  • Utilize complianceDIN EN ISO 2178The standard electromagnetic field induction method can measure the thickness of coatings such as zinc, chromium, copper, paint, varnish, fibers, etc. on steel and iron materials (the instrument can automatically identify the metal substrate material and select the appropriate measurement method)

characteristic:

One handed operation, easy to use: simply place the instrument on the workpiece to see the measurement reading

Ergonomically designed double-sided backlit display panel, with readings visible from any measurement position

Built in dual function measuring probe, suspended spring, maintaining constant pressure and low contact pressure

Automatic switch function

Measurement can be carried out without calibration

Quick selection of functions through an easy-to-use menu

Can set upper and lower limits

If the measurement result exceeds or meets the upper and lower limit settings, the instrument will emit different sounds and flashing lights as prompts

Automatic selection of measurement method, electromagnetic field induction or eddy current response

When there is no battery, it can still be stored1000A data

Radio transmission reading (required)RS232Receiver) directly toEXCELForm (requires software cooperation)PC-DATEX)If you want better acceptance ability, the receiver can useRS232External wires on the computer

Statistical evaluation

Free choice of deviation compensation value

Correction: Standard film or substrate can be used for correction

Technical specifications:

measuring range NF,Iso/Fe:0-2000μm(80mils)
Iso/NF:0-2000μm(80mils)
measurement accuracy (Using Fischer standard film as a criterion) 0-50μm(0-2mils),±1μm(0.04mils)
50-1000μm(2-40mils),±2%
1000-2000 μ m (40-80mils), maximum 3% higher
Communication interface Radio bandwidth transmission 868MHz (Europe)/915MHz (United States)
weight 60g (2.1 amps) without battery included
volume W×D×H = 64mm × 30mm × 85mm
Battery 2×LR6.AA
Protection standards IP 52
Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!