Main features
Infrared calibration transfer source, standard transfer radiation pyrometer
• Transfer source DG10Ncal
Calibration temperature range: 300~1800 ° C
• Wavelength 1.5~1.8 μ m
The German DIAS transmission source DG10Ncal has high accuracy and is commonly used as a calibration source for infrared thermometers and thermal imagers.
Performance
Digital monochrome infrared thermometer, temperature measurement range: 300~1800 ° C
The response time is fixed at 1 second
• Minimum measurable diameter of 3mm @ 600mm
Communication interface RS485
On site display and on-site button adjustment parameters
• Lens aiming
Technical performance
model |
DG10Ncal |
Measuring range |
300 ~1800 °C |
spectral range |
1.5 μm~1.8 μm |
LENS |
Fixed focal length, minimum measurement target 3mm@600mm |
Distance coefficient ratio |
200:1 |
temperature measurement error |
Refer to the factory inspection form |
repeatability |
0.1K(Tamb= 23 °C, ε = 1, t95 = 1s) |
NETD |
0.1K(TU= 23 °C, ε = 1, t95 = 1s) |
Response time (t)95) |
Fixed at 1 second |
emissivity |
0.900~1.100, adjustable through communication interface or infrared thermometer button |
output signal |
0/4 ~ 20 mA, Adjustable through software, temperature linear, maximum load: 500 Ω |
communication interface |
Connect via the included USB cable |
sighting system |
Aim through a lens |
Adjustable parameters |
Adjustable through software: SSE correction (emissivity), temperature unit in ° C or ° F |
power supply |
Wall mounted 100~240VAC |
power consumption |
Maximum 1.5 W |
operate temperature |
23 °C ± 3 K |
storage temperature |
0~70 °C |
weight |
Approximately 520 grams |
size |
54 × 54 mm, length 170 mm |
protection grade |
IP65 compliant with DIN EN 60529 and DIN 40050 |
test specification |
EN 55 011:1998, class A |
CE certification |
Compliant with EU regulations (EN 50 011) |
scope of supply |
PYROSPOT DG10Ncal, Installation base with tripod adapter, operation manual, 10 point inspection sheet (traceable to PTB standard), software PYROSOFT Spot (no connection cable, please order separately), radiation source size effect (SSE) curve, various globally compatible plug adapters (EU/US/UK/AU) USB connection cables, plastic packaging box |
Ultra high precision
DS10Ncal_DG10Ncal | ||
Radiation temperature ° C |
Actual measurement error ° C |
% |
299.95 |
0.21 |
0.07% |
450.009 |
0.078 |
0.02% |
600.026 |
0.079 |
0.01% |
750.02 |
0.12 |
0.02% |
899.97 |
0.15 |
0.02% |
1050.1 |
0.83 |
0.08% |
1197.27 |
0.89 |
0.07% |
1397.5 |
1 |
0.07% |
1796.9 |
1.2 |
0.07% |
List of Black body Furnace Models:
Infrared standard transfer source Transfer Pyrometer for blackbody furnace