Application areas:
Widely used in fields such as plant physiology, agriculture, horticulture, forestry, microbiology, algae biology, life sciences, marine biology, zoology, human medicine, and environmental science.
1. Measure the photosynthetic rate of plant leaves, algae, and green tissues. Especially suitable for determining the photosynthetic rate of mosses, Arabidopsis thaliana, fruit peels, and plants with acid metabolism that cannot be measured by a photosynthesis analyzer
2. Measure the changes in respiratory rate and pathways of animals, plant tissue cells, and microorganisms, and analyze the changes in cyanide resistant respiratory pathways, cytochrome oxidase pathways, glycolysis pathways, and tricarboxylic acid pathways
3. Measurement of mitochondrial respiration in animals, plants, and other organismsIStateIIStateIIIStateIVState breathing,Research on respiratory control rate andP/Ocompare
4. Determine the enzymatic reaction process involving aerobic participation. Such as polyphenol oxidase, lipoxygenaseH2O2Enzyme and other activities
5. Determine the oxygen release rate of chemically synthesized oxygen releasing substances
Main functions:
l Highly integrated controller. No need to connect to a computer for operation, system settings and data recording are all controlled by the controller4Execute an operation key
l LCDThe settings of the display screen display system and the size of the oxygen signal
l The system has a small size and is easy to carry outdoors for operation. Can be used in an environment without AC power12VBattery as a power source
Origin and Manufacturer: UKHanshatech
Main technical indicators:
1. host Machine: Highly integrated controller, integrated magnetic stirrer, speed250~900 rpmadjustable
2. Sample dosage:0.2~2.5 ml
3. Measurement range:0~100% O2
4. Oxygen resolution:20At ℃10×10-6 μ mol· ml
5. Oxygen consumption: Oxygen consumption<0.015 μ mol · ht-1
6. light Source: White Light Source: High Intensity3000μ mol· m-2·s-1, adjustable light intensity
7. light/Temperature gauge: Temperature measurement range:0~50℃Light intensity measurement range:0~5000 μ mol· m-2·s-1
8. Temperature control: The sample reaction chamber has a circulating water bath interlayer, which can be connected to an external circulating water bath for temperature control
9. gain/Compensation control: signal compensation1 mVIn units, the signal gain is measured by2In units
10. Signal gain mode: coarse adjustment gain: ×1, ×2, ×5, ×10, ×20, ×50, ×100
Fine tuning gain:1 mVIn units
11. Electrode output:21% O2 When is1 μA;10%~90%
12. Response time:10%~90%Within the scope<5second
13. Polarization voltage:0.4~0.9V, default polarization voltage0.7V
14. Working voltage:95~260V