BINDER FD115 drying oven
1、 Product Introduction:
1. Temperature range: ambient temperature plus 10 ℃ to 300 ℃
2. Up to 30% more energy-efficient than common devices on the market
3. Utilizing APT.line ™ Technology achieves high temperature accuracy
4. Circulating air
5. Controller with LCD display
6. Exhaust valve electromechanical control
7. 2 chrome plated sockets
8. Stackable
9. Integrated independently adjustable temperature safety device level 2 (DIN 12880) with optical alarm
10. Ergonomic handle design
11. USB interface for recording data
2、 Technical parameters:
PRODUCT MODEL | FD115-230V1 | FD115UL-120V1 |
Choose a model | standard | standard |
Item code | 9010-0305 | 9010-0306 |
measure | ||
Internal volume [L] | 116 | 116 |
Net weight of equipment (unloaded) [kg] | 50 | 50 |
Maximum total load [kg] | 150 | 150 |
Maximum load per partition [kg] | 30 | 30 |
The dimensions of the casing do not include additional components and connectors | ||
Net weight width [mm] | 710 | 710 |
Net weight at height [mm] | 735 | 735 |
Depth net weight [mm] | 605 | 605 |
Wall distance from rear [mm] | 160 | 160 |
Wall to side distance [mm] | 100 | 100 |
Internal dimensions | ||
Width [mm] | 550 | 550 |
Height [mm] | 550 | 550 |
Depth [mm] | 385 | 385 |
The number of doors | ||
Outer door | 1 | 1 |
Data related to the environment | ||
Energy consumption at 150 ℃ [Wh/h] | 340 | 340 |
Sound pressure level [dB (A)] | 43 | 43 |
firmware | ||
Number of partitions (hours/maximum) | 2/5 | 2/5 |
Temperature performance data | ||
The temperature range exceeds the indoor temperature by 10 ℃ to [℃] | 300 | 300 |
Temperature deviation is 150 ℃ [± K] | 1.7 | 1.7 |
Temperature fluctuation at 150 ℃ [± K] | 0.3 | 0.3 |
Heating time to 150 ℃ [min] | 18 | 18 |
The recovery time after opening the door in 30 seconds is 150 ℃ [min] | 4 | 4 |
Electric data | ||
Rated voltage [V] | 230 | 120 |
Power frequency [Hz] | 50/60 | 60 |
Rated power [kW] | 1.3 | 1.4 |
Equipment safety device [A] | 6.3 | 12.5 |
Phase (rated voltage) | 1~ | 1~ |
1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. The technical data assumes a fan speed of 100%. We reserve the right to make technical changes.