Shanghai Huanao Technology Co., Ltd
Home>Products>BINDER ED720 drying oven
BINDER ED720 drying oven
BINDER ED720 drying oven
Product details

BINDER ED720 drying oven

1、 Product Introduction:

1. Temperature range: ambient temperature plus 5 ℃ to 300 ℃

2. Up to 30% more energy-efficient than common devices on the market

3. Utilizing APT.line ™ Technology achieves high temperature accuracy

4. Natural convection

5. Controller with LCD display

6. Exhaust valve electromechanical control

7. 2 chrome plated sockets

8. Integrated independently adjustable temperature safety device level 2 (DIN 12880) with optical alarm

9. Ergonomic handle design

10. USB interface for recording data

2、 Technical parameters:

PRODUCT MODEL ED720-400V1
Choose a model standard
Item code 9010-0341
measure
Internal volume [L] 743
Net weight of equipment (unloaded) [kg] 169
Maximum total load [kg] 315
Maximum load per partition [kg] 45
The dimensions of the casing do not include additional components and connectors
Net weight width [mm] 1165
Net weight at height [mm] 1590
Depth net weight [mm] 870
Wall to side distance [mm] 100
Wall distance from rear [mm] 160
Internal dimensions
Width [mm] 960
Height [mm] 1280
Depth [mm] 605
The number of doors
Inner door 2
Outer door 2
Data related to the environment
Energy consumption at 150 ℃ [Wh/h] 700
firmware
Number of partitions (hours/maximum) 2/16
Temperature performance data
The temperature range exceeds the indoor temperature by 5 ℃ to [℃] 300
Temperature deviation is 150 ℃ [± K] 3.2
Temperature fluctuation at 150 ℃ [± K] 0.8
Heating time to 150 ℃ [min] 85
The recovery time after opening the door in 30 seconds is 150 ℃ [min] 25
Electric data
Rated voltage [V] 400
Power frequency [Hz] 50/60
Rated power [kW] 4.1
Equipment safety device [A] 16
Phase (rated voltage) 3~

1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. We reserve the right to make technical changes.