Nanjing Kelunde Scientific Instrument Co., Ltd
Home>Products>Automated Combination System - Semi automatic 894 Professional CVS
Automated Combination System - Semi automatic 894 Professional CVS
The semi-automatic professional 894 Professional CVS is a convenient analytical instrument for determining organic additives in electroplating tanks u
Product details

Order number: 2.894.1210

Semi automatic professional type894 Professional CVSIt is used«Cyclic voltammetry dissolution method»CVS)The«Cyclic pulse voltammetry dissolution method»CPVS)The chronopotentiometric method(CP)A convenient analytical instrument for determining organic additives in electroplating tanks. This confirmed Wan Tong electrode technology and newly designed potential/Constant potential meter and external high-performancevivaSoftware integration, inCVSIt presents new prospects. With the help of replaceable measuring heads, quick switching between various applications using different electrodes can be achieved. The potentiostat with a certified calibrator automatically flushes for calibration before each measurement to ensure possible high accuracy. The integrated temperature measurement input can monitor the temperature of the solution during the measurement process.

Two accompanying800 DosinoAuxiliary solutions can be automatically added during the measurement process, such asVMSStandard solution or used for dilution titration(DT)The sample.

To carry out control, data collection, and evaluation work, it is necessaryvivaSoftware.

Semi automatic professional type894 Professional CVSA large number of accessories and measuring heads for rotating disc electrodes were included in the supply. Electrode group andvivaThe license must be ordered separately.

technical parameter

Size remarks

No attachments

Size, unit:mm(Wide)/high/Deep)

188/294/406

technical parameter

Potentiostat

·Maximum output voltage:±25 V

·Maximum output current:± 224 mA

·Scanning voltage range:±5 V

·Resolution of measured current (within the minimum measurement range):6 fA

Current measurement technology

·CVS«Cyclic voltammetry dissolution method»

·CPVS«Cyclic pulse voltammetry dissolution method»

·CP(Timing potential method, constant current and measurement of open circuit point)

Correction technology

·DT(Dilution titration method)

·LAT(Linear approximation method)

·MLAT(Improved Linear Approximation Method)

·RC(Response curve)

Working electrode

·Rotating disk electrode(Rotating disk electrodesRDE)The material is glassy carbon(Glassy Carbon)The“Ultra Trace”(Ultra trace) graphitePt(Platinum)Ag(Silver)Au(Gold)

·revolutions:100 To3000 min-1

Reference electrode

·Ag/AgCl/KCl 3 mol/L

Auxiliary electrode

·Pt(Platinum)

·Glassy carbon(Glassy CarbonGC

port/Interface

·Built inUSBInterface

·MSBInterface, used to connect up to four devices800 DosinoOr other flying accessories

·Measurement input terminal, used for temperature measurement(Pt1000