QualitybyDesign
Optimized design that meets both your budget planning and water quality requirements
Introduction: Aquaplore2S is an optimized design designed for users without special environmental management requirements. Adopting various special purification technologies to reduce overall purification costs, coupled with our latest intelligent management system, we monitor all water production processes in real-time to ensure that the quality of your ultrapure water is not affected The system also supports cloud maintenance services, making your management work easier. You can easily enjoy reliable ultrapure water quality without having to pay expensive procurement and maintenance costs.
| Dimensions and Weight | |||
| size | H x W x D; 580x 395x 535 mm | ||
| packing weight | 38kg | ||
| Water inlet requirements | |||
| water source | Tap water inflow | ||
| traffic | ≥ 50 L/h | ||
| Conductivity (25 ° C) | < 2000 μS/cm | ||
| pressure | 1-5 Kg | ||
| temperature | 5-35 °C | ||
| free chlorine | < 1 ppm | ||
| PH value | 6.5-8.5 | ||
| Total hardness (as CaCO)3Calculation) | <200ppm | ||
| Electricity requirements | |||
| Electricity input | 100-240 VAC,50-60 Hz | ||
| system voltage | 24V DC | ||
| System power | 221W(max) | ||
| Noise level | < 50 dB | ||
Laboratory Ultra Pure Water Series Aquaplore 2S
Quality by Design
Optimized design that meets both your budget planning and water quality requirements
Unique DP-RO reverse osmosis technology
The patented DP-RO deep desalination technology can improve the overall desalination rate of the system by more than 99.5%, reduce the impact of changes in raw water quality, and stably produce third grade water that meets ISO standards under the condition of raw water conductivity not exceeding 1000 μ S/cm. At the same time, it greatly improves the durability of the system, extends the service life of the ultra pure water filter, and reduces the production cost of ultra pure water, solving the high purchase and frequent maintenance costs brought by continuous electric desalination technology.
Compared with electrolytic deionization technology (EDI), DP-RO can quickly purify multiple batches of compliant pure water without the need for a large buffer tank, making it ready to use and more suitable for laboratory environments with frequent water production and extraction, and reducing the risk of microbial contamination caused by long-term storage.
The most complete process monitoring
The most comprehensive monitoring range, from pre-treatment, reverse osmosis, water storage tanks, ultra purification, post-treatment to water intake units, can be monitored in real-time without any omissions.
The most monitored items include water quality, pressure, water temperature, flow rate, water storage capacity, pure water production capacity, pure/ultrapure water intake capacity, system operation time period, intake time period, intake personnel, cumulative intake capacity, remaining consumables purification capacity, etc. All key parameters are comprehensively and timely monitored and self diagnosed online to ensure the normal operation of the system. And store all operational data in a timely manner for future queries.
Clear icon touch display
The intuitive icon interface allows you to clearly access all system information displayed on the screen, from operational status to water quality; From purification ability to water usage record. All operations can be easily done with just a few touches, simple and easy to learn.
Easy and timely cloud maintenance
You can now focus more on your work. Our cloud service can provide independent databases for each end user and create files and analyze relevant data for easy user queries. As long as you download your operating data to USB and upload it to our cloud service center, well-trained and service certified engineers will perform data detection and analysis on your operating data, and provide the most professional preventive maintenance advice and implementation plan, avoiding you from occupying your own work time to manage the ultrapure water system, while ensuring that your experimental water meets regulatory requirements.
| Product Series | A2S-05-OR/A2S-10-OR | A2S-05-CE/A2S-10-CE | A2S-05-BE/A2S-10-BE |
| water quality parameters | |||
| ultrapure waterElectrical resistivity (25 ° C) | 18.2MΩ•cm | 18.2MΩ•cm | 18.2MΩ•cm |
| Pure water conductivity * (25 ° C) | 1-5 μS/cm | 1-5 μS/cm | 1-5 μS/cm |
| Total Organic Carbon (TOC) * * (treated with DP-RO) | - | < 5ppb | <1-3ppb |
| bacteria | - | < 0.1 CFU/ml | < 0.1 CFU/ml |
| bacterial endotoxin | - | - | < 0.001 EU/ml |
| Particulate matter (>0.2um) | <1/ml | <1/ml | <1/ml |
| Inorganic ion removal rate | ≥99.5% | ≥99.5% | ≥99.5% |
| heavy metal ions | ≤ 0.1 ppb | ≤ 0.1 ppb | ≤ 0.1 ppb |
| Silicon removal rate | > 99% | > 99% | > 99% |
| Water quantity parameters | |||
| Secondary reverse osmosis water production (25 ° C) | 5、10L/H | ||
| Water intake for pure water/ultrapure water respectively | 10ml-99L, The value can be set arbitrarily, with an accuracy of ± 2ml within 500ml and ± 1% above 500ml | ||
| Water intake flow rate | 1-2L/min | ||
| Ultra pure aquatic water quantity*** | 6000L-10.000L | ||
| Water utilization rate | 60-80% | 60-80% | 60-80% |
| Purification technology | |||
| Tap Water Pre treatment | yes | yes | yes |
| Reverse osmosis (DP-RO) | yes | yes | yes |
| Ultra purification column | yes | yes | yes |
| Photo oxidation | No | 185/254nm UV | 185/254nm UV |
| Ultrafiltration column | No | No | 6KD |
| 0.2um Terminal Filter (UF) | yes | yes | yes |
| *Under the condition of general municipal water supply and conductivity<1000 μ S/cm | |||
| **Under the condition of general municipal water supply and raw water TOC<1000ppb | |||
| ***Different depending on the inflow situation | |||