product details
SidePakTM AM520 Individual Exposure Dust AnalyzerIt is a compact, portable, battery powered light scattering laser photometer with data recording function, which can provide real-time aerosol mass concentration readings of dust, smoke, and fog in the worker's breathing area. The newly designed particle size cutting sampling head improves the performance of the equipment in measuring mass concentration and can measure different particle size cutting points for PM 10, respiratory dust (PM 4), PM 5 (Chinese respiratory dust), PM 2.5, PM 1, and diesel engine emission particles (DPM) PM 0.8. This monitor is a solution for real-time individual aerosol sampling in various workplace environments, including general industry, foundry industry, construction sites, chemical plants, refineries, petrochemical industry, power and public facilities, transportation, aerospace, maritime, enclosed spaces, mining industry, etc.
New features
·Robust particle size cutter for higher concentration range - *!
·Sound and light alarm - users can receive visual and auditory reminders!
·0.8 μ m diesel engine emission particle size cutter - same type *!
·PM 5 Chinese respiratory dust particle size cutter
·High capacity battery - with a continuous running time of over 20 hours, ensuring its flexibility in use!
·Color OLED display screen
Features and advantages
·Real time quality concentration measurement and data recording function for 'on-site' data analysis
·Button menu operation or through improved TrakPro ™ Programming operations using v5 software
·User selectable alarm level to alert workers to high aerosol concentration exposure levels
·The newly designed sturdy particle size cutter enables the instrument to provide reliable particle size measurements at higher aerosol concentrations and longer operating times
Industrial applications
·Aerospace
·Automotive industry
·Chemical industry
·Construction industry
·General Industry
·Medical Health
·Manufacturing industry
·Mining industry
·Petrochemical industry
·Pharmaceutical industry
·Power generation industry
·Transportation industry
Product parameters:
sensitivity |
sensor type |
90 ° light scattering, 650nm laser diode |
calibration |
According to ISO12103-1 standard, use Al to test the respiratory dust particle size range of dust, and calibrate using the sampling and weighing reference method |
|
aerosol |
concentration range |
0.001-100mg/m3 |
particle size range |
0.1-10μm |
|
resolution |
0.001mg/m3 |
|
Zeroing stability |
Using a 10 second time constant, the change over 24 hours is 0.001mg/m3 of soil |
|
temperature coefficient |
Approximately+0.0005mg/m/℃ (Change in temperature during one zeroing on the instrument) |
|
internet traffic |
range |
User adjustable range: 0-1.BL/min Note: The higher range is connected to SidePak AM520 Determine the particle size cutter and cyclone separator of the monitoring device. |
temperature range |
Operation range |
32°F-120°F(0°C-50°C) |
Storage range |
-4°F-140°F(-20°C-60°C) |
|
Operating humidity |
0-95% relative humidity, no condensation |
|
time constant (OLED display) |
range |
User adjustable range: 1 second, 5 seconds, 10 seconds, 15 seconds, and 30 seconds |
Built in sampling head |
Standard sampling head, no specific particle size range |
PM1.0 particle size cutter intercepts 50% at 1.0 µ m |
PM2.5 particle size cutter intercepts 50% at 2.5 µ m | ||
PM5.0 particle size cutter intercepts 50% at 5.0 µ m | ||
PM10 particle size cutter intercepts 50% at 10.0 µ m | ||
Detachable cyclone separator |
4 µ m Dorr Oliver intercepts 50% at 4.0 µ m |
|
0.8 µ m DPM intercepts 50% at 0.8 µ m | ||
alarm |
Alert Type |
Universal, S |
default setting |
0.5mg/m3, 1mg/m3, 3mg/m3, 5mg/m3 15mg/m3, closed Range programmed through TrakPro software: 0.001-999mg/m3 |
|
alarm indicator |
95 decibel speaker, flashing red LED, Flashing red OLEO display |
|
Second call delay |
30 seconds, 1 minute, 3 minutes, 5 minutes, 10 minutes, off |
|
data record |
data point |
Approximately 80000 (recorded once per minute for 55 days) |
Record Interval |
User adjustable intervals: 1 second, 1 minute, 5 minutes, 15 minutes, 30 minutes |
|
User selectable correction factors |
Factory settings |
1.0 (user adjustable) |
Environment settings |
0.38 (user adjustable) |
|
User defined settings |
5 groups (user-defined tags through TrakPro software) |
|
physical property |
0.1-10.0 (user adjustable) |
|
physical device |
External dimensions |
5.1 x 3.7 x 3.1 inches 129.5 mm x 94 mm x 78.4 mm Containing 803300 batteries |
weight |
22 ounces (0.62 kg) containing 803300 batteries |
|
display |
160 x 128 resolution color OLED display |
|
Tripod installation nut |
¼ -20 internal thread |
|
battery |
Battery capacity |
5400 mAH 捚 ion battery pack |
run time |
Using PM2.5 cutter at 1.7Lpm for over 20 hours |
|
charging time |
4.0 hours (typical case of completely depleted battery) |
|
power supply (pin 803302) |
Input Voltage |
100-240V AC, 50-60Hz |
Output voltage |
12V DC @ 3.0A |
|
communication interface |
type |
US82.0 |
Instrument connector |
USB Micro-8 (socket) |
|
maintenance |
Manufacturer cleaning/calibration |
Suggest conducting it every year |
User zero calibration |
Before each use |
|
User traffic calibration |
According to the requirements of the sampling probe |
|
authentication |
AM520 using TSI battery pack P/N 803300 |
|
exempt |
EN61326-1:2013 |
|
emission |
EN61326-1:2013 Class B |
|
secure |
IEC 61010-1:2010 |
|
IEC 60825-1:2014 |
Contains components
·Soft single portable bag
·100-240VAC power supply
·USB data cable
·Impact particle size cutter assembly
·Dorr Oliver cyclone separator component
·Diesel engine particulate matter cyclone separator component
·Screwdriver
·TrakPro Data Analysis Software CD
·Micro USB connection cable
·Zero filter for calibration
·Sampling tube
·Calibration certificate
·User Guide
·Quick reference card
Standard configuration list:
Serial Number |
Instrument Name |
model |
quantity |
notes |
1 |
AM520 Individual Exposure Dust Analyzer |
AM520 |
1 unit |
|
packing list | ||||
Accessory Name |
quantity |
|||
Soft single portable bag |
1 piece |
|||
100-240VAC power supply |
1 piece |
|||
USB data cable |
1 piece |
|||
Impact particle size cutter assembly |
1 set |
|||
Dorr Oliver cyclone separator component |
1 piece |
|||
Diesel engine particulate matter cyclone separator component |
1 piece |
|||
screwdriver |
1 piece |
|||
TrakPro Data Analysis Software CD |
1 copy |
|||
Micro USB connection cable |
1 piece |
|||
Zero filter for calibration |
1 piece |
|||
Sampling tube |
1 piece |
|||
CALIBRATION CERTIFICATE |
1 sheet |
|||
user's guide |
1 book |
|||
Quick reference card |
1 sheet |